Beyond INF-B Betreibershandbuch Seite 70

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 132
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 69
16
Input Configuration Menu
60
16.2 Ad.RAtE Enter ANALOG TO DIGITAL RATE SUBMENU:
Press ‘RESET/ENTER 7) Display flashes "0" or previous selection.
Press ‘/MAX 8) Scroll through the 6 available selections as 0, 1, 2, 3, 4, and 5 for
your choice of selection. Refer to below Table for your specific application
conversion speed. Only Rate 0 and 1 can reject 60 Hz / 50 Hz noise.
Refer to Table 33-1 in Specification Section for different input and outputs types.
Rate 0 1 2 3 4 5
Time for 60hz(ms) 134.9 68.2 34.9 18.2 9.9 5.7
Time for 50Hz(ms) 161.9 81.9 41.9 21.9 11.9 6.9
Table 16-1. A/D (Analog to Digital) Rate Option Table
Press ‘RESET/ENTER 9) Display advances to "MOdE" Input Mode Configuration Submenu. If a
change was made, the controller will also show "StOREd" stored
message momentarily.
16.3 MOdE Enter INPUT MODE CONFIGURATION SUBMENU:
This submenu is only accessible if "tC" Thermocouple or "bRIdGE"
Bridge is your selection of input type.
Seitenansicht 69
1 2 ... 65 66 67 68 69 70 71 72 73 74 75 ... 131 132

Kommentare zu diesen Handbüchern

Keine Kommentare